Accuracy:
Better than ± 5 % of reading or ± 1 mg/L whichever is greater
Analog Outputs:
Two 4-20 mA outputs suitable for recorders or PID control. Output span programmable over any portion of the 1 to 500 mg/L range (130 Vac isolation from earth ground)
Calibration Interval:
60 minutes
Certifications:
NRTL certified to UL and CSA standards and CE approved
Cycle Time:
8 minutes
Data Communications Distance:
Maximum node-to-node distance is 400 m (1312'). Maximum total wire length is 500 m (1640')
Depth:
(21")
Dimensions (H x W x D):
522 mm x 627 mm x 526 mm
Drain:
Gravity, air break or vent recommended
Drain Fitting:
1/4" OD NPT barbed hose fitting
Enclosure Rating:
NEMA 4X (indoor)/IEC 529 (IP66) with provision for air purge. Reagent enclosure is drip-proof.
Height :
(21")
Inlet:
Sample filter: 3/4" NPT male or female
Mounting:
Wall, bench and panel
Parameter:
Alkalinity
Power requirements (Hz):
50/60 Hz
Power requirements (Voltage):
90 - 240 V AC
Range:
1 - 1000 mg/L total alkalinity as CaCO3
5 - 1000 mg/L CaCO3 as phenolphthalein alkalinity
Relay output:
Two SPDT relays with contacts rated for 5A resistive load at 230 Vac. Additional relays available through optional Signal Output Module.
Repeatability:
Better than 3 % of reading or ± 0.6 mg/L, whichever is greater
Response time:
Less than 10 minutes for 90% response to setup change at instrument sample fitting
Sample Flow Rate:
20 - 1000 mL/min
Sample Interval:
Optionally capable of analyzing up to three sample streams
Sample Pressure:
0.03 - 6.9 bar
Sample Temperature:
5 - 50 °C (41 - 122 °F)
Weight:
25.5 kg
What's included?:
APA6000 Alkalinity analyzer, installation kit, one month's supply of reagents, a maintenance kit, Hach user manual.
Width:
(25")