Chemistries INQUIRY: +91 9916040249 - e-Shop: +91 63668 87212

CLF10 sc,free Cl,CL cell NPT

CLF10 sc,free Cl,CL cell NPT
Product #: LXV45A.98.12022
Unit Price Contact Hach

Compatible with Hach's Plug and Play Digital Controllers

The CLF10 sc and CLT10 sc analyzers leverage Hach's exclusive self diagnostics to alert users when the process has changed or the instrument needs servicing. Diagnostic features include the Cal Watch algorithm for warning of pH and chlorine calibration deviation and a non-contacting flow sensor for notification of insufficient sample flow.

No Reagent Replacement, No Waste Stream

Chlorine measurement with an amperometric analyzer, such as the CLF10 sc or CLT10 sc, does not require reagents, eliminating the need for routine reagent replacement and waste stream management.

EPA Compliant

The CLF10 sc and CLT10 sc analyzers allow for real-time control of disinfection processes by providing continuous readings that indicate when treatment conditions have changed.

Advanced Diagnostics

The CLF10 sc and CLT10 sc analyzers can be used with any Hach sc digital controller. Whether you're measuring turbidity or chlorine, you only need to learn one controller for all your water analysis measurement points. Hach sc controllers have no complicated wiring or setup procedures. Just plug in any Hach digital sensor and it's ready to use without software configuration.

Real-time Process Monitoring and Control

In accordance with EPA Method 334.0, the CLF10 sc and CLT10 sc analyzers can be used for reporting chlorine residual measurements. Additionally, Hach has created a suite of laboratory products and methods to help with startup and quality control procedures required in Method 334.0. (See page 4 for a partial listing of accessories.)

What's included?

CLF10sc Free Chlorine NPT Cell

Specifications

Accuracy: Free Chlorine: ± 3 % of the reference test** (DPD) at constant pH less than 7.2 (±0.2 pH unit)
Accuracy 2: Free Chlorine: ± 10 % of the reference test** (DPD) at stable pH less than 8.5 (±0.5 pH unit from the pH at calibration)
Accuracy 3: Total Chlorine: ± 10 % of the reference test** (DPD) at stable pH less than 8.5 (±0.5 pH unit from the pH at calibration)
Accuracy 4: Total Chlorine: ± 20 % of the reference test** (DPD) at stable pH greater than 8.5
Cable Connection: 5 pin, M12 connector
Cable Length: 1 m (between gateway and sc controller)
Calibration Method: 1-point or 2-point (zero and slope) calibration
Certifications: CE compliant for conducted and radiated emissions:
- CISPR 11 (Class A limits)
- EMC Immunity EN 61326-1 (Industrial limits)
Additional certifications when connected to an sc controller
Connection Drain Line: 1/2 in ID
Connections: 1/4 in OD
Controller Compatibility: sc controller platform
Description : CLF10sc Free Chlorine, NPT Cell
Diameter: 25 mm
Dimensions: mm x 25 mm (7.68 in x 0.98 in)
  Sensor Only: 195
Dimensions (H x W x D): 483 mm x 495 mm x 151 mm
Includes: CLF10sc Free Chlorine NPT Cell
Interferences: Free chlorine: monochloramine, chlorine dioxide, ozone, and chalk deposits.
Total chlorine: chlorine dioxide, ozone and chalk deposits
Length: Sensor: 195 mm
Lower Limit of Detection (LOD): 30 ppb (0.03 ppm)
Material: Panel: Corrosion-resistant, fully-submersible (stainless steel, PVC, silicon rubber and polycarbonate)
Measurement method: Reagentless, electrochemical, three-electrode amperometric system
Mounting: Panel or Wall Mounting
Operating temperature range: 0 - 45 °C
Parameter: Free Chlorine
pH Range: 4 - 9
Power requirements (Voltage): 12 V DC
Pressure Range: 0.5 bar, no pressure impulses and/or vibrations
Range: 0 - 20 ppm Cl₂
Repeatability: 30 ppb or 3%, whichever is greater
Response time: Free Chlorine: 140 s or less for 90% change (T90)
at a stable temperature and pH
Response Time: Total Chlorine: 100 s or less for 90% change (T90)
at a stable temperature and pH
Sample Flow Rate: 30 - 50 L/h (7.9 - 13.2 gal/hour), Optimal is 40 L/hour (10.5 gal/hour)
Sample Temperature: 5 - 45°C (41 - 113°F)
Storage conditions: -20 °C to 60 °C dry
Temperature Compensation: Internal temperature compensation in the sensor
  Internal temperature compensation in the sensor
Warranty: 12 months
Weight: 5.5 kg Approximately 5.5 kg (12 lbs)
Panel and empty panel-mounted components only
What's included?: CLF10sc Free Chlorine NPT Cell
Width Panel: 495.3 mm