Accuracy:
± 0.1 % of span, at 25 °C
Calibration Method:
2-point Buffer (for pH only)
1-point Sample (for pH or ORP)
Communication:
0/4 to 20 mA with 0.004 mA (12-bit) resolution (during calibration, the analog output is automatically held at the last measured value and, upon completion, returned to its active state)
Dimensions (H x W x D):
95 mm x 95 mm x 60 mm
Display:
2-line by 16 character LCD
Load:
Maximum permissible load dependent on power supply voltage, transmitter hookup arrangement, and wire resistance
Material:
Enclosure: Polycarbonate
Measuring Range ORP:
-2100 - 2100 mV
Measuring Range pH:
Selectable Ranges:
-2.0 - 14.0 pH
-2.00 - 14.00 pH
Measuring Range Temperature:
-20 to 200 °C
Memory Backup:
All user settings are retained indefinitely without battery backup
Mounting:
Panel, wall, pipe, or integral sensor mounting
Operating temperature range:
-20 - 60 °C at 0 - 95% RH (non-condensing)
Parameter:
pH/ORP
Power requirements (Voltage):
12 - 30 V DC
Relative Humidity:
0 - 95 % non-condensing
Repeatability:
± 0.05 % of span, at 25 °C
Response Time T90:
1 - 60 s with output filter setting of zero
Sensitivity:
± 0.05 % of span, at 25 °C
Special Notes:
The measured pH (or ORP) and temperature can be displayed separately or together on one screen. Both analog output values are
shown together on one screen
Storage conditions:
-30 to 70 °C 0-95% Relative humidity, non-condensing
Temperature Compensation:
-10 - 110 °C
Automatic with selection for NTC 300 Ω, Pt1000 or Pt100 or manually fixed at a user-entered temperature; additional selectable temperature correction factors (ammonia, morpholine or user-set pH/°C linear slope)
Temperature Drift:
0.02 % of span per °C (zero and span), at 25 °C
Warranty:
12 months
Weight:
0.280 kg
What's included?:
Controller, User Manual
Width:
with back cover