Accuracy:
±25% of set point value
Alarm:
2 consectutive cycles above set point activate alarm. 1 cycle below set point cancels alarm.
Alarm points:
2 mg/L of CaCO3 trip point
Certifications:
NRTL certified to UL and CSA standards, and CE approved
Cycle Time:
2 average
Dimensions (H x W x D):
419 mm x 318 mm x 178 mm
Fitting type:
Inlet: ¼-in. OD tube quick-disconnect fitting
Outlet: ½-in. ID flexible hose
Maintenance Interval:
Standardise and replenish reagents every 2 months using EDTA or high-level sample
Replace pump tubes every 6 months
Material Enclosures:
ABS plastic, large plastic windows to view alarm and reagents level
Measurement method:
Colorimetric
Operating Humidity:
5 - 95%, non-condensing
Operating temperature range:
5 - 40 °C
Outputs:
1 SPDT relay
Power requirements (Hz):
50/60 Hz
Power requirements (Voltage):
115/230 V AC
Reagent Consumption:
500 mL each indicator and buffer every two months
Reagent Requirements:
each of indicator and buffer. Replenish every two months for continuous operation.
Repeatability:
± 10 % of set point value in 0.3 - 2 mg/L ranges
± 4 % of set point value in 5 - 100 mg/L ranges
Sample Flow Rate:
50 - 500 mL/min
Sample Pressure:
0.07 - 0.34 bar (0.10 bar is optimum)
Sample conditioning: 0.10 - 5.17 bar
Storage conditions:
-40 - 60 °C
Warranty:
12 months
Weight:
11.3 kg
What's included?:
SP510 Hardness Monitor, two month's supply of reagents, an installation kit, a maintenance kit, Hach user manual.