Promotions INQUIRY: +91 9916040249

Amtax sc Ammonium Analyzer, Outdoor, 0.05-20.0mg/L NH₄-N, Two Channel Continuous Sample, 115-230 VAC

Amtax sc Ammonium Analyzer, Outdoor, 0.05-20.0mg/L NH₄-N, Two Channel Continuous Sample, 115-230 VAC
Product #: 6157500
Unit Price Contact Hach

Wide Measurement Range

Amtax sc Ammonium Analyzers are available in four measuring ranges. With detection limits as low as 0.02 mg/L and ranges up to 1000 mg/L, the Amtax sc can be used anywhere in the wastewater treatment process, from the influent or start of the nitrification process with high NH4-N levels to the effluent with only low ammonium levels. In the lowest range it can also be used for drinking water applications.

Fast Response Time for Control of Nutrient Removal Process

Nutrient removal is a critical and sensitive process in wastewater treatment. Quick response is vital for optimal operation. The fast response time of only 5 minutesand selection of the measurement interval, make the Amtax sc Ammonium Analyzer the fastest responding ammonia analyzer on the market.

Easy Installation

The Amtax sc Analyzer is designed to be installed at the measuring location, including outdoor locations. The housing is weatherproof and lockable so that the analyzer can be installed right at the basin, even in the toughest climates.
The Amtax sc easily mounts to a wall, rail- and stand-mounting options are available. Due to its SC technology the analyzer is ready for use in the network of the SC controller ("plug-and-play").

Low Maintenance and Simple Operation

The analyzer calibrates and cleans itself automatically and is equipped with comprehensive self-diagnostic routines. With at least a 3-month supply of reagents (for 5 min interval), the Amtax sc analyzer offers the most economical reagent consumption. Reagents and wear parts are easily accessible.

Optional with Sample Preparation or Second Channel

The Amtax sc analyzer works with a low-maintenance filtration system (e.g. Filtrax) that removes particles larger than 0.15 µm. The filtrate is sent to the analyzer while air bubbles constantly clean the filtration membrane.
The analyzer is also available as two channel version for measurement at two different sampling points.

What's included?

0.05 - 20 mg/L NH₄-N, 115 VAC Amtax sc 2-Channel Ammonium Analyzer, 3 month supply of all necessary consumables at 5 minute cycle time, user manuals.

Specifications

Accuracy: Using standard solutions: 3% + 0.05 mg/L
Cable Length: 2 m fixed data cable at analyzer
  2 m fixed power cable at analyzer
Data Transfer: Current output, relays and bus communication via SC controller
Dimensions (H x W x D): 720 mm x 540 mm x 390 mm
Enclosure Rating: IP55
Filtersystem included: No, use e.g. Filtrax
Flow: 1 - 20 L/h Sample (free of suspended solids)
Fuse Protection: Via SC1000 controller or power box. Maximum of 2 analysers per SC1000 or power box
Indoor/Outdoor: Outdoor
Lower Limit of Detection (LOD): 0.05 mg/L NH4-N
  1.0 mg/L NH4–N
Material: ASA/PC UV-resistant
Measurement method: GSE (Gas Sensitive Electrode) with screw-on membrane cap
Measuring Interval: 5 - 120 min, adjustable per 5 min.
Measuring Principle: NH4 is being converted to NH3 gas, wich is being measured at the GSE.
Number of Channels: 2
Operating Humidity: 95% relative humidity, non-condensing.
Operating temperature range: -20 - 45 °C
Outputs: Current output, relays and bus communication via SC controller
Parameter: Ammonium
Permissible Chloride Range: Max. Cl- concentration: 1000 mg/L
pH Range: 5 - 9 pH
Power requirements (Hz): 50 Hz
Power requirements (Voltage): 115 - 230 V AC, powered by Power Box or SC Controller
Pressure Range: -30 - 50 mbar With continuous sample preparation; at overflow vessel
Range: 0.05 - 20.0 mg/L NH₄-N
Reproducibility: 2% + 0.05 mg/L
Response Time: < 5 minutes
Sample Level: Sample level in basin and filtration system must be below analyzer.
Sample Quality: Ultra filtrated or comparable.
Sample Temperature: 4 - 40 °C (39 - 104 °F)
Sampling Option: two channel continuous sample from filtration unit, e.g. FILTRAX
Special Feature: Lockable housing; automatic cleaning and calibration; extensive self-diagnostics
Storage conditions: 4 to 55 °C less than 95% Relative Humidity
Warranty: 12 months
Weight: Without sample preparation system and without chemicals: 31 kg
What's included?: 0.05 - 20 mg/L NH₄-N, 115 VAC Amtax sc 2-Channel Ammonium Analyzer, 3 month supply of all necessary consumables at 5 minute cycle time, user manuals.