Chemistries INQUIRY: +91 9916040249 - e-Shop: +91 63668 87212

ANATEL TOC600, TOC Analyzer

ANATEL TOC600, TOC Analyzer
Product #: FG1100600
Unit Price Contact Hach

Calibration and validation

The TOC600 Analyzer can be calibrated with from one ot three TOC standards using pre-packaged Calibration Standards available from Hach Ultra. Calibration and instrument performance can be validated online using Hach Ultra Validation Kit.

USP complinace for conductivity and system suitability

The TOC Analyzer's limit response and response efficiency are calculated online automatically to determine complinace with USP TOC Method <643>. The cell constant also can be verified online to comply with USP Conductivity Method <645>.

Hot water application

Fully compatible with water temperatures up to 70°C. For water temperatures from 70°C to 95°C, an external heat exchanger is required. The only wetted part in the Anatel Heat Exchanger is 316 stainless steel tubing.

Individual alarm monitoring

TOC levels may be monitored for all connected TOC600 Analyzers and compared to user-defined alarm limits for each instrument. The user is alerted to any limit excursions by LED. An audible beep or user-supplied external alarm may be enabled for additional warning. The TOC600 Analyzer can also be set to signal at USP recommended uncompensated conductivity levels.

Local Area Networking (LAN) capability

The ability to link as many as eight TOC600 Analyzers to a single C80 Controller provides timely readings of TOC levels even at the most distant and crucial areas of the water system. The C80 Controller can be mounted wherever convenient to make the TOC levels at several key locations immediately available on a single display. Analog, Digital, I/O, Serial and 12 VDC Bias connections allow quick and easy connection to any external device to further enhance system capabilities.

What's included?

Includes: TOC600 Analyzer, power cord, network tee and terminator, inlet filter, sample and drain tubing, tube and fitting, operator manual and Standard Operating Procedures (SOPs).

Specifications

What's included?: Includes: TOC600 Analyzer, power cord, network tee and terminator, inlet filter, sample and drain tubing, tube and fitting, operator manual and Standard Operating Procedures (SOPs).