Simple preparation

TNTplus vials use Dosicaps - freeze-dried reagents integrated into sealed cap - that are easier to use than Powder Pillows or Liquid Reagents, without any risk of contamination. The boxes and vials are color-coded for a fast and easy parameter and range recognition of exactly the test you need.
Step-by-step illustrated test methods are printed on the box as a quick reference and can also be called up in the instrument menu.

Comprehensive documentation

Measurement results are documented on the detailed level with timestamp, operator ID, absorbance reading, and calculated concentration. The 2D barcode delivers the lot number and expiry date, logged with every result.

Fast execution

A 2D barcode on the LCK cuvettes is automatically read by the Hach DR spectrophotometer to identify the appropriate method and take the measurement.
The vial spins to take 10-fold absorbance readings that will be averaged for result determination to exclude scratches and fingerprints. Instrument calibration verification and high instrument stability all combine to eliminate the need to run reagent blanks.

Customisable

With the ability to store hundreds of user-determined methods, operators are able to tailor the DR3900 to meet the everyday needs of the plant.
Being able to optimise and customise the method portfolio, combined with regular software updates and Claros connectivity, makes the DR3900 the ultimate solution to water quality lab needs.

What's included?

Instruments include: Adapter A for 1 in. round and 1 cm square cells, matched pair of 1 in. square glass sample cells, light shield, dust cover, printed basic user manual and benchtop power supply with 115 and 230V power cords.

Specifications

Beam Height: 10 mm
Data Logger: 2000 measured values (Result, Date, Time, Sample ID, User ID)
Dimensions (H x W x D): 151 mm x 350 mm x 255 mm
Display: 7" TFT
Display Resolution: WVGA (800 pix x 480 pix)
Display Size: 7 inch (17.8 cm)
Display Type: Coloured touch-screen
Enclosure Rating: IP30
Interfaces: USB type A (2)
USB type B
Ethernet
Manual Languages: German
English
French
Italian
Spanish
Portuguese (PT)
Czech
Danish
Dutch
Hungarian
Polish
Romanian
Russian
Slovenian
Swedish
Turkish
Greek
Finnish
Croatian
Bulgarian
Serbian
Slovakian
Max. operating humidity: 80 %
Max. Storage Humidity: 80 %
Operating Conditions: 10 - 40 °C
Operating Mode: Transmittance (%), Absorbance and Concentration, Scanning
Optical System: Reference beam, spectral
Photometric Accuracy: 5 mAbs @ 0.0 - 0.5 Abs
Photometric Accuracy 2: 1 % at 0.50 - 2.0 Abs
Photometric Linearity: < 0.5 % - 2 Abs
Photometric Linearity 2: ≤ 1 % at >2 Abs with neutral glass at 546 nm
Photometric Measuring Range: ± 3.0 Abs (wavelength range 340 - 900 nm)
Power Requirements: With external power supply
Power requirements (Hz): 50/60 Hz
Power requirements (Voltage): 110 - 240 V AC
Power Supply: Desk Power Supply
Preprogrammed Methods: > 220
Sample Cell Compatibility: Rectangular: 10 mm, 50 mm, 1 inch
Round: 13 mm, 1 inch
Source Lamp: Gas-filled Tungsten (visible)
Specific Technology: RFID not applicable with this model
Spectral Bandwidth: 5 nm
Standard accessories: None
Storage conditions: -30 - 60 °C
Stray Light: < 0.1 % T at 340 nm with NaNO2
User Interface Languages: Bulgarian, Chinese, Croatian, Czech, Danish,
Dutch, English, Finnish, French, German,
Greek, Hungarian, Italian, Japanese, Korean,
Polish, Portuguese - Brasilian, Potuguese,
Russian, Serbian, Slovakian, Slowenian,
Spanish, Swedish, Turkish
User Programs: 100
Warranty: 12 months
Wavelength Accuracy: ± 1.5 nm (wavelength range 340 - 900 nm)
Wavelength Calibration: automatic
Wavelength Range: 320 - 1100 nm
Wavelength Reproducibility: ± 0.1 nm
Wavelength Resolution: 1 nm
Wavelength Selection: Automatic
Weight: 4.2 kg
What's included?: Instruments include: Adapter A for 1 in. round and 1 cm square cells, matched pair of 1 in. square glass sample cells, light shield, dust cover, printed basic user manual and benchtop power supply with 115 and 230V power cords.