EZ2304 Total Nickel + Ni(II) Analyser

EZ2304 Total Nickel + Ni(II) Analyser
Product #: EZ2304.XXXXXXXX
Unit Price Contact Hach

Specifications

Alarm: 1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Ambient Temperature: 10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Analog Outputs: Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
Calibration: Automatic, 2-point; frequency freely programmable
Certifications: CE compliant / ETL certified
Cycle Time: 30 minutes (dilution + 10 min.)
Demineralised water: For rinsing / dilution
Digital outputs: Optional: Modbus (TCP/IP, RS485)
Dimensions (H x W x D): 690 mm x 465 mm x 330 mm
Drain: Atmospheric pressure, vented, min. Ø 32 mm
Earth connection: Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Instrument air: Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Interferences: Metal ions like Aluminium Al(III), Bismuth Bi(II), Cadmium Cd(II), Chromium Cr(III), Cobalt Co(II), Copper Cu(II), Iron Fe(II), Iron Fe(III), Lead Pb(II), Manganese Mn(II), Magnesium Mg(II), Mercury Hg(II), Palladium Pd(II), Platinum Pt(II), Silver Ag(I), Tin Sn(II), Zinc Zn(II). Large amounts of colour and turbidity interferes. Fats, oil, proteins, surfactants and tar.
Lower Limit of Detection (LOD): ≤ 10 µg/L
Material: Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
Measurement method: Colorimetric measurement using DMG colour solution
Model: EZ2304
Number of sample streams: 1, 2, 4 or 8
Parameter: Nickel, total; Nickel Ni(II), dissolved
Power: 230 VAC, 50/60 Hz
120 VAC, 50/60 Hz
Max. power consumption: 440 VA
Precision: Better than 2% full scale range for standard test solutions
Protection Class: Protection class: Analyser cabinet: IP44 / Panel PC: IP65
Range: 20 - 500 µg/L Ni
  20 - 500 µg/L Ni

Optional:
10 - 125 µg/L
10 - 250 µg/L
160 - 2000 µg/L
320 - 5000 µg/L
Reagent Requirements: Keep between 10 - 30 °C
Sample Flow Rate: 100 - 300 mL/min
Sample Pressure: By external overflow vessel
Sample Quality: Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Sample Temperature: 10 - 30 °C
Validation: Automatic; frequency freely programmable
Warranty: 12 months
Weight: 25 kg