Accuracy:
Turbidity up to 1000 NTU:
without calibration < 5% of the measured value ±0.01 NTU
with calibration < 1% of the measured value ±0.01 NTU
Automatic Wiper:
Yes
Cable Length:
10 m (optional extension cables available)
Calibration Method:
Turbidity: Formazin or STABLCAL Standard (at 800 NTU).
Suspended Solids: Sample specific, based on gravimetric TSS analysis with a correction factor procedure.
Certifications:
CE
Controller:
Sensor only
Diameter Sensor:
60 mm
Flow:
Max. 3 m/s (the presence of air bubbles affects the measurement)
Length:
200 mm
Maintenance Interval:
1 h/month
Material:
Optics Carrier and Sleeve: stainless steel 1.4571; Wiper Arm: stainless steel 1.4581; Wiper Rubber: silicone rubber (standard) Optional: FKM/FPM (LZX578); Wiper Shaft: stainless steel 1.4571
Max. cable length Sensor-to-Controller:
100 m
Measuring Principle:
Infrared Duo scattered light technique for color-independent turbidity measurement Turbidity in accordance with DIN EN ISO 7027 / TS equivalent DIN 38414
Measuring Range TSS:
0.001 mg/L - 50 g/L
Measuring Range Turbidity:
0.001 - 4000 NTU
Model:
Solitax ts-line sc
Mounting:
Immersion Immersion
Operating temperature range:
0 - 40 °C
Parameter:
Suspended Solids, Turbidity
Pressure Range:
< 6 bar
Repeatability:
< 3 %
Repeatability Turbidity:
< 1 %
Response Time T90:
1 - 300 s adjustable
Warranty:
12 months
Weight Sensor:
1.8 kg
What's included?:
Solitax ts-Line Turbidity Sensor, User Manual, Backwash Filter Mounting Kit